Model Number 71674035
Device Problem Contamination (1120)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2020
Event Type  malfunction  
Event Description
It was reported that, during surgery, the nurse found an unidentified material on the device.No delay reported.No patient injuries reported.An s&n backup was not required.
 
Manufacturer Narrative
The device, used in treatment was not returned for evaluation, reporting event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.For cleaning procedures please refer to smith and nephew¿s recommended cleaning methods given in our cleaning and sterilization brochure-¿instructions for care, maintenance, cleaning and sterilization of smith and nephew orthopedic devices¿, this device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPRESSION SCREW HEXDRIVER
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9861019
MDR Text Key184316593
Report Number1020279-2020-00978
Device Sequence Number1
Product Code HXX
UDI-Device Identifier03596010560551
UDI-Public03596010560551
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674035
Device Catalogue Number71674035
Device Lot Number09BM08285
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/20/2020
Initial Date FDA Received03/20/2020
Supplement Dates Manufacturer Received05/31/2020
Supplement Dates FDA Received06/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-