• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT4
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.Product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported that during a cataract extraction with intraocular lens (iol) implant procedure, that there was a defect on the iol before inserting the iol into the cartridge.The surgeon used a backup iol.There was no risk to the patient.No further information is available.
 
Manufacturer Narrative
Product evaluation: the product was returned for analysis and damage was observed to the lens.Additional observations were as follows: iol returned positioned incorrectly in the iol case and is pressed down into well area of iol case base, resulting in deformation of the iol.Solution is dried on the iol.One haptic is broken/torn and is returned.The optic is scratched/marked rejectable.We are unable to determine the root cause for the reported complaint "defect on the lens".The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on the iol.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol contributed to the event.All product and batch history records are quality reviewed prior to product release.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
MDR Report Key9862339
MDR Text Key186734335
Report Number9612169-2020-00108
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberSN6AT4
Device Catalogue NumberSN6AT4.100
Device Lot Number21145055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Initial Date Manufacturer Received 02/25/2020
Initial Date FDA Received03/20/2020
Supplement Dates Manufacturer Received07/16/2020
Supplement Dates FDA Received08/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-