MEDTRONIC NAVIGATION, INC (LITTLETON) O1; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number 9733346
Device Problem Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer representative went to the site to test the imaging system.The reported issue was confirmed and the bulkhead on the back of the mobile view station (mvs) was replaced and secured with the additional cable provided with the mvs ethernet cable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used in a sacroiliac and thoracolumbar procedure.It was reported that the communication between the imaging system and the navigation system was not solid.It was only solid when the ethernet cable was pushed and maintained straight on the back of the mobile view station (mvs).There was less than an hour delay in the case.No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O1
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9865332
MDR Text Key184480585
Report Number3004785967-2020-00388
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733346
Device Catalogue Number9733346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2020
Initial Date FDA Received03/23/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age21 YR
Patient Weight58
-
-