DRÄGERWERK AG & CO. KGAA FABIUS GS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8607400
Device Problems Device Reprocessing Problem (1091); Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/04/2020
Event Type  malfunction  
Manufacturer Narrative
The electronic log file was available for investigation.Several log entries indicate that there was a negative pressure detected various times.Based on these log entries it was concluded that the auxiliary air valve was (temporarily) stuck open, leading to restricted automatic ventilation.Depending on the actual situation the system will post corresponding alarms, such as insp o2 low; minute volume low; insp pres not reach; apnea pressure.Software performs a valve test as part of the leak test.The number of similar cases is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that there was failure of the ventilator during use.There was no injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS GS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
MDR Report Key9870869
MDR Text Key184630164
Report Number9611500-2020-00089
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011404
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8607400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2020
Initial Date FDA Received03/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2006
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse