• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN KNEE FEMORAL ADAPTOR BOLT; UNK KNEE FEMORAL ADAPTOR BOLT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN KNEE FEMORAL ADAPTOR BOLT; UNK KNEE FEMORAL ADAPTOR BOLT Back to Search Results
Catalog Number UNK KNEE FEMORAL ADAPTOR BOLT
Device Problems Fracture (1260); Naturally Worn (2988)
Patient Problems Foreign Body Reaction (1868); Not Applicable (3189); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This complaint is to capture the multiple cases with the same problem reported in (b)(4).Tc3 bolt and adaptor broke in patient.Outcome was a revison for tc3 femoral component.I don¿t have any implant product lots or part numbers but see descriptions below: femur: 5 tc3.Distal augments: 16 medial and lateral.+2 bolt & 5 degree adaptor.Sz 5 ps rp poly which was removed by osteotome poly from poly bullet.There was metal debris in the joint and the sleeve was well fixed.Xrays and implant will be uploaded.Surgeon dr.(b)(6) removed the femoral component after 30 minutes of work.The 5 degree adaptor was still engaged in the sleeve which we were able to break the taper with slap hammer and five grip using the bolt.There have been multiple cases with the same problem which i have not been a part of.Surgeon was able to re-engage morse taper with an srom hinge implant.Patient status/ outcome / consequences --> yes, patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)?--> revison, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, if yes, describe --> xrays will be attached , is the patient part of a clinical study --> unknown, (b)(4).Device property of -->none, device in possession of -->none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN KNEE FEMORAL ADAPTOR BOLT
Type of Device
UNK KNEE FEMORAL ADAPTOR BOLT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9874441
MDR Text Key189845427
Report Number1818910-2020-09132
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK KNEE FEMORAL ADAPTOR BOLT
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/27/2020
Initial Date FDA Received03/24/2020
Supplement Dates Manufacturer Received04/16/2020
Supplement Dates FDA Received04/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN KNEE FEMORAL ADAPTOR; UNKNOWN KNEE FEMORAL ADAPTOR BOLT
Patient Outcome(s) Required Intervention;
-
-