• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY MAILLEFER X-SMART ENDO MOTOR WITH CONTRA ANGLE HANDPIECE; HANDPIECE, DIRECT DRIVE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY MAILLEFER X-SMART ENDO MOTOR WITH CONTRA ANGLE HANDPIECE; HANDPIECE, DIRECT DRIVE, AC-POWERED Back to Search Results
Catalog Number A100400000000
Device Problems Defective Device (2588); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a x-smart contra angle won't hold files; no injury resulted.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the device for evaluation.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART ENDO MOTOR WITH CONTRA ANGLE HANDPIECE
Type of Device
HANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
MDR Report Key9879399
MDR Text Key186734039
Report Number8031010-2020-00044
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
PMA/PMN Number
K990682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA100400000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/24/2020
Initial Date FDA Received03/25/2020
Supplement Dates Manufacturer Received07/07/2020
Supplement Dates FDA Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-