Model Number 2545-00-547
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirms the complaint; the articulation surface has fractured across the location peg.Heavy gouging was also noted.Examination of the returned device confirmed the reported event.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the surgeon mangled sz 7 attune ps artic surf with osteotome and broke shim in half.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP PS ARTIC SURF SZ7
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopedics drive
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopedics drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9884368
MDR Text Key185148166
Report Number1818910-2020-09342
Device Sequence Number1
Product Code JWH
UDI-Device Identifier10603295132875
UDI-Public10603295132875
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2545-00-547
Device Catalogue Number254500547
Device Lot NumberMVMBSZ050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2020
Initial Date FDA Received03/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-