Model Number DB-1200-S
Device Problem Unexpected Therapeutic Results (1631)
Patient Problems Undesired Nerve Stimulation (1980); Electric Shock (2554); No Code Available (3191)
Event Date 03/04/2020
Event Type  Injury  
Event Description
It was reported that the during a reprogramming visit the physician made several changes to the patients medication and referred him to physical therapy.A cat scan was taken that indicated no changes to the location of the leads.It was later reported that when the patient was trying to connect their ipg the patient felt multiple adverse effect such as severe shock, tingling and muscle tightening in his face, right side of the body, arm and leg which only lasted for a few minutes.During a follow up visit with the physician the patient experienced the same adverse effects while connecting to the clinician programmer but did not last as long.The patient was reprogrammed and the patients adverse symptoms did not occur again.Two database analysis were performed and revealed no anomalies, the ipg appears to be working as expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GEVIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key9886950
MDR Text Key185915218
Report Number3006630150-2020-01477
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729984443
UDI-Public08714729984443
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/10/2021
Device Model NumberDB-1200-S
Device Catalogue NumberDB-1200-S
Device Lot Number739832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/09/2020
Initial Date FDA Received03/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
-
-