• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT266
Device Problems Fitting Problem (2183); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The three complaint rt266 infant dual heated evaqua2 breathing circuits are currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the connector of three rt266 infant dual heated evaqua2 breathing circuits do not fit.There was no reported patient involvement.
 
Manufacturer Narrative
Ps335381.Additional information was received from the customer noting five to ten circuits were affected.Method: the five to ten complaint rt266 infant dual heated evaqua2 breathing circuits were not returned to f&p for evaluation.Our investigation is thus based on the information and photograph provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph did not indicate any faults with the complaint device.Conclusion: the provided photograph did not indicate any faults with the complaint device.However, without the complaint device, we are unable to identify the cause of the reported event.All rt266 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject infant breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: - "check all connections are tight before use." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "set appropriate ventilator alarms.' - "visually inspect breathing sets for damage (e.G.A crushed tube or cracked connector) before use and replace if damaged." - "appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm, or death.".
 
Event Description
A healthcare facility in sweden reported via a fisher & paykel healthcare (f&p) field representative that the connector of three rt266 infant dual heated evaqua2 breathing circuits do not fit.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
MDR Report Key9896269
MDR Text Key187428812
Report Number9611451-2020-00335
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number2100924174
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/01/2020
Initial Date FDA Received03/29/2020
Supplement Dates Manufacturer Received06/21/2020
Supplement Dates FDA Received07/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-