DATEX-OHMEDA, INC. AVANCE CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 03/04/2020
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The switch was replaced, and the unit was returned to service.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported the bag to vent switch cracked resulting in an inability to switch ventilation modes as expected.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
MDR Report Key9904294
MDR Text Key222169304
Report Number2112667-2020-01105
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K123125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2020
Initial Date FDA Received03/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-