Model Number CLV-190
Device Problems Fire (1245); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The subject device has not returned for evaluation.As stated on the reported event, the cleaning protocol of the user was changed immediately when the issue happened and to date there was no patient/user harm or injury was reported due to the event.If additional or new information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported that the user had sprayed alcohol in the socket of a clv-190 scope which caused fire.It was reported to be a quick flash of fire from an alcohol aerosol in which it went out/disappeared quickly on its own.The issue occurred when the cleaning protocol was being performed in between patient procedure.According to the report the cleaning protocol was changed immediately when the issue happened.This reported incident according to the reporter happened seven years ago and the reported statement from the reporter was made when troubleshooting a video issue during a facility service visit at the unrelated site.There was no patient or user harm and injury reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9904453
MDR Text Key185749122
Report Number8010047-2020-01905
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/06/2020
Initial Date FDA Received03/31/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-