• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND GLENOSPHERE ECC D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND GLENOSPHERE ECC D42MM; DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE Back to Search Results
Model Number 130760042
Device Problem Difficult to Insert (1316)
Patient Problems Not Applicable (3189); No Code Available (3191)
Event Date 03/12/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Intraoperatively during a primary shoulder-tep it was not possible to connect the metaglene with the glenosphere.Surgical delay less than 10 min., no adverse patient consequences.Procedure was finishes successfully as another metaglene plus glenosphere implant was available.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) used to capture the surgical intervention and surgery prolonged.Product complaint #
=
> pc-(b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no additional related incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE ECC D42MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9911451
MDR Text Key190973144
Report Number1818910-2020-09903
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027751
UDI-Public10603295027751
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760042
Device Catalogue Number130760042
Device Lot Number5359796
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/16/2020
Initial Date FDA Received04/01/2020
Supplement Dates Manufacturer Received03/23/2020
06/08/2020
Supplement Dates FDA Received04/15/2020
06/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DXTEND METAGLENE
-
-