• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AHCT
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Date 03/12/2020
Event Type  malfunction  
Manufacturer Narrative
Evaluation is in progress, but not yet concluded the reported complaint was observed by comparing the readings with another bpm, as well as using external thermometer.The manufacturer's subsidiary's service technician did not find any errors by the first inspection, only a 'f00d' message.
 
Event Description
The service technician reported that during routine testing of the device at the manufacturer's subsidiary's service center, the blood parameter monitor (bpm) temperature measurements were constantly fluctuating and slightly lower than the actual temperature of the fluid.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the arterial blood parameter monitor (bpm) to differ considerably from room temperature as read with a multimeter.It was determined that the arterial bpm was the cause of the inaccuracy.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9922392
MDR Text Key195888551
Report Number1828100-2020-00151
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00886799001622
UDI-Public(01)00886799001622(11)170107
Combination Product (y/n)N
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/12/2020
Initial Date FDA Received04/03/2020
Supplement Dates Manufacturer Received04/16/2020
Supplement Dates FDA Received04/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-