Model Number 0168L12 |
Device Problems
Inflation Problem (1310); Failure to Infuse (2340)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
It was reported that it was difficult to inflate during pretest.
|
|
Manufacturer Narrative
|
The reported event was confirmed as design-related.An amber lubricath foley was returned without its original packaging.The balloon was attempted to be inflated with 10cc of water using a leur lock syringe.The inflation funnel began to balloon instead of infusing into the balloon.The catheter valve was cut to find the occlusion area.Water was attempted to be infused using a slip tip syringe with no success.The funnel was cut closer to the shaft and water was able to be infused using the same syringe.No occlusion material was seen.The issue appears to be isolated to upper portion of the inflation funnel.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "-warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.-valve type: use luer slip tip syringe.Do not use needle.-recommended inflation capacities:5cc balloon: use 10ml sterile water.-visually inspect the product for any imperfections or surface deterioration prior to use.".
|
|
Event Description
|
It was reported that it was difficult to inflate during pretest.Per additional information from the investigator via email 23may2020; upon evaluation of the sample it was found that the inner lumen of the inflation funnel was collapsed.
|
|
Search Alerts/Recalls
|