• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ1 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ1 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-12-070
Device Problem Degraded (1153)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191)
Event Date 04/16/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that patient had increasing hip & groin pain.Head & liner exchange.After review of medical records, it was indicated that patient presents with progressive weight-bearing pain and some intermittent mechanical symptoms of catching or ratcheting in her right hip.Her evaluation revealed that she did have elevated cobalt and chromium ion levels, and an mri did show a moderate-sized periarticular fluid collection.The patient was then revised for failed right total hip arthroplasty due to metallosis and local adverse tissue reaction as well as early periacetabular osteolysis.Operative notes reported that there was blackish and grayish metallosis staining within the synovial lining diffusely of the hip joint, and there was some wear debris within the capsular structures.There was clear trunnionosis with wear debris at the junction of the femoral head and the trunnion.There was some mild yellowish wear debris on the posterior aspect of the liner, surgeon felt the majority of the metallosis wear debris was from the trunnion.There was evidence for some osteolysis.Doi: (b)(6) 2007; dor: (b)(6) 2019; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ1 HI OFF
Type of Device
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9928393
MDR Text Key186547231
Report Number1818910-2020-10231
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295060123
UDI-Public10603295060123
Combination Product (y/n)N
PMA/PMN Number
K011489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1570-12-070
Device Catalogue Number157012070
Device Lot NumberBY1A81000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/11/2020
Initial Date FDA Received04/06/2020
Supplement Dates Manufacturer Received10/19/2021
Supplement Dates FDA Received10/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient Weight63
-
-