Involved components nj015t- bicontact n plasmapore 8/10 size 15mm- 51508869a.Nj102- biolox prosthesis head 8/10 28mm m- 51523514.General information we received a complaint about a nh474 - sc/msc pe-insert 28mm 56/58 asym.Regarding a revision surgery 10 years and 7 months post-operatively.We did not receive the components for investigation.Consequences for the patient post-operative medical intervention was necessary à revision surgery.Failure description due to a lack of components, a failure description is not possible.Investigation due to a lack of components, an investigation is not possible.Batch history review the device history records have been checked for the available lot numbers and found to be according to the specification, valid at the time of production.No further complaints registered against the same lot number of the pe-insert (51477463).Conclusion and root cause the failure is most probably usage / patient related.Rationale based on the information provided and without the product for investigation a clear conclusion can not be drawn.There is no indication for a material or manufacturing failure on the basis of the dhr review.As mentioned by the surgeon (see figure 1) the occurred dislocations may be traced back to the implantation situation and the cup angle, therefore we assess this as the most probable root cause.
|