Model Number SCORPION-MULTIFIRE NEEDLE
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 03/17/2020
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event Description
It was reported during a rotator cuff repair procedure the tip of the ar-13995n, multifire scorpion needle broke off after one singular fire.The rep reported there was no abnormal tissue.The needle tip was lodged into the cuff tissue and unable to be recovered.Additional information has been requested.Additional information received on 04/03/2020: the sales rep reported the surgeon spent over 20 minutes trying to recover the broken needle tip, including x-ray use.No unplanned incisions were necessary.A second ar-13995n from the same lot was used to complete the procedure with a full rotator cuff repair using normal technique.The complaint device was discarded and will not be returning for evaluation.The rep reported at this time the surgeon does not plan to perform a second surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPION-MULTIFIRE NEEDLE
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key9935342
MDR Text Key186809704
Report Number1220246-2020-01779
Device Sequence Number1
Product Code MDM
UDI-Device Identifier00888867015951
UDI-Public00888867015951
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberSCORPION-MULTIFIRE NEEDLE
Device Catalogue NumberAR-13995N
Device Lot Number10516579
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/17/2020
Initial Date FDA Received04/08/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/18/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-