• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 811474
Device Problem Decoupling (1145)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/13/2020
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the centrifugal head became decoupled from the centrifugal motor.The end user noticed decoupling, reattached the centrifugal pump and the decoupling was resolved.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Per clinical review: on march 13, 2020, during cardiopulmonary bypass (cpb), the perfusion team saw that the sp45 had decoupled from the drive motor on the heart lung machine (hlm).The system was set up and primed without issue for the procedure on (b)(6) 2020.The flow and revolutions per minute (rpm) during the procedure were appropriate, but the end user noticed during the cpb that the connection was not complete on the drive motor, as stated by the end user 'noticed the decoupling, then reattached the centrifugal pump'.It is not known what preceded the end user noticing the issue, but patient forward flow was consistent.The end user was able to recouple (properly attach) the centrifugal disposable to the drive motor without issue, and without a delay in the continuation of the surgical procedure.There was no harm or blood loss associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9936238
MDR Text Key195913966
Report Number1828100-2020-00158
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number811474
Device Catalogue Number811474
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2020
Initial Date FDA Received04/08/2020
Supplement Dates Manufacturer Received04/17/2020
Supplement Dates FDA Received04/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-