Catalog Number AM-05501
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 03/31/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
The report states: anesthesia technician notified me this morning of reports from the ob director of 2 kits from same lot with broken vials.
 
Manufacturer Narrative
Qn#(b)(4).A device history record review was performed based on potential lot numbers (23f19d0087 and 23f19d0007) provided by the customer.Lot number 23f19d0087 is for kit cdc-45703-xp1a so the device history record review was performed for lot number 23f19d0007.A device history record review was performed with a potentially relevant finding.A nonconformance was initiated only as a general nc to temporarily allow certain finished good part numbers to be packaged with additional gauze in order to better secure medication ampules in the inner tray.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed based on a potential lot number provided by the customer.A device history record review was performed with a potentially relevant finding.Based on the information provided, the potential root cause of this issue is design related.A capa was initiated to further investigate this complaint issue.
 
Event Description
The report states: anesthesia technician notified me this morning of reports from the ob director of 2 kits from same lot with broken vials.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9936501
MDR Text Key186866561
Report Number1036844-2020-00137
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAM-05501
Device Lot Number23F19D0087
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/31/2020
Initial Date FDA Received04/08/2020
Supplement Dates Manufacturer Received05/15/2020
Supplement Dates FDA Received05/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-