Model Number 80440 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not find a conclusive cause for the higher-than-expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.Investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: w035420018871 the platelet collection set is not available for return because it was discarded by the customer.
|
|
Manufacturer Narrative
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process.A follow up report will be provided.
|
|
Manufacturer Narrative
|
This report is being filed to provide corrected information in b.5 and b.6 investigation is in process.A follow-up report will be provided.
|
|
Event Description
|
Corrected donor unit #: (b)(4).
|
|
Manufacturer Narrative
|
Root cause: the analysis of the run data file did not find a conclusive cause for the higher-than- expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher- than- expected wbc content in the platelet product could be donor-related.
|
|