COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Failure to Advance (2524)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Pma/510(k) #: exempt.Occupation: equipment and product standardisation nurse (skin and wound) ¿ supply chain management.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Event Description
|
It was reported an unknown number of ultrathane mac-loc locking loop multipurpose drainage catheters were unable to be inserted during nephrostomy tube changes.Additional information regarding the event, quantity of devices, failure mode, and patient outcome has been requested but is currently unavailable.
|
|
Event Description
|
See h10.
|
|
Manufacturer Narrative
|
Correction: this report is being sent to indicate the complaint event is not reportable.Upon further investigation, this event is not reportable.The customer previously indicated that the catheter was unable to be inserted.Clarification was received (b)(6) 2020 stating that it was the flexible stiffener that could not be advanced into the catheter.There is no information confirming device malfunction or serious injury.Replacement of device to complete procedure is negligible harm and does not meet the criteria for a reportable event.A review of risk documentation does not indicate that this event is likely to cause serious injury if it were to reoccur.As there are no recorded incidences of serious injury due to the complaint event, and it is not likely that serious injury would result if the event were to recur, per 21cfr part 803.50 the complaint event is considered not reportable.
|
|
Search Alerts/Recalls
|
|
|