FISHER & PAYKEL HEALTHCARE LTD ICON SERIES CPAP; BZD Back to Search Results
Model Number ICONPBN
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint icon cpap was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer.Results: customer reported that the icon cpap has a frayed power cord.Conclusion: based on the information we are unable to determine what caused the reported fault.During initial assembly of the icon cpap, all power cords are visually inspected for damage.Once the assembly process is completed, all icon cpap devices are visually inspected again before release for distribution.This suggests the damage occured after it had been distributed.The icon cpap is designed to the electrical safety standards, ul60601-1 and as/nzs 3200.1.The materials used in the thermoplastic components of the mains connector and the cases are flame retardant according to ul 60601-1 and as/nzs 3200.1.Our user instructions that accompany the icon state the following: "only operate if the device, power cord and plug are dry and in good working order." "do not operate the device, water chamber or breathing tube if it is dropped, damaged or not working as intended.".
 
Event Description
A healthcare facility in (b)(6) reported that an icon cpap humidifier had a frayed power cord.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICON SERIES CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology dr., suite 100
irvine, ca 
4534000142
MDR Report Key9960967
MDR Text Key199861759
Report Number9611451-2020-00376
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K094040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberICONPBN
Device Catalogue NumberICONPBN
Device Lot Number140827
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/17/2020
Initial Date FDA Received04/14/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-