Model Number SPSCAA
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint f&p sleepstyle series cpap was received at fisher & paykel healthcare service center in (b)(6) and was visually inspected.Results: visual inspection confirmed that one pin had detached from the mains inlet socket.Conclusion: the reported incident was traced to an issue in the assembly process of the supplied mains inlet connector component.The supplier of the component was notified and they have made changes to the assembly process.As part of our ongoing product improvement initiatives, we recently implemented a gauge test which identifies and rejects any potentially faulty mains inlet sockets prior to assembly into the sleepstyle.The subject sleepstyle was manufactured prior to implementation of these measures.Our user instructions that accompany the f&p sleepstyle state the following: do not use if the device, power cord or accessories are damaged, deformed, or cracked.Do not pull on the power cord as it may become damaged.Turn the device off at the power supply, then remove the power cord from the rear of the device.
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the f&p sleepstyle auto cpap is damaged.Upon device assessment it was observed that the main power input socket was dislodged.There was no patient involvement and no user consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F&P SLEEPSTYLE AUTO CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, ca 
7923912142
MDR Report Key9966291
MDR Text Key199861375
Report Number9611451-2020-00385
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K173193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSPSCAA
Device Catalogue NumberSPSCAA
Device Lot Number2100443870
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/17/2020
Initial Date FDA Received04/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-