Model Number SA60AT
Device Problems Degraded (1153); Material Opacification (1426); Structural Problem (2506); Unintended Movement (3026)
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Manufacturer Narrative
A sample product was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
In a literature study entitled, "outcome of various hydrophobic acrylic intraocular lens implantations in children with congenital cataract," the performance of different intraocular lenses (iols) after cataract surgery in children aged less than two years with regards to rates of visual axis opacification were evaluated.For the 16 children associated with the lens model of this report (11 male & 5 female), the complications encountered during the study were documented as pupil capture (1) , pigment on the iol (3), lens decentered (1), lens shifted location in the eye (1), pupillary membrane (5), membranectomy (5) and posterior capsular opacification (4).In conclusion, the study indicated that the opacifications evaluated were comparable for all lens models studied.This report is associated with the events for the first indicated lens in the study charts.This is one of two medical device reports associated with the article.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key9979018
MDR Text Key188326196
Report Number1119421-2020-00704
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.XXX
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/30/2020
Initial Date FDA Received04/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-