• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BUNNELL, INC. LIFEPULSE; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BUNNELL, INC. LIFEPULSE; VENTILATOR, HIGH FREQUENCY Back to Search Results
Device Problem Pressure Problem (3012)
Patient Problems Bradycardia (1751); Low Oxygen Saturation (2477)
Event Date 04/22/2019
Event Type  malfunction  
Event Description
Servo pressure on jet ventilator decreased significantly on two separate occasions, causing the patient to brady and desat.Company called and suggested to replace patient box.Box was replaced.After the investigation, it was noted the pressures decreased on the jet ventilator which the respiratory therapist (rt) noticed.The patient was taken off the vent and neo puffed.The rt called the support line for the (b)(6) company.Through the discussion, it was determined to change the patient box out.This was completed and the pressures on the jet came back into the ordered parameters.The defective patient box is being sent back to the (b)(6) company.No harm was caused to the patient from this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPULSE
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
BUNNELL, INC.
436 lawndale dr.
s salt lake UT 84115
MDR Report Key9980817
MDR Text Key188345118
Report Number9980817
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/17/2020
Date Report to Manufacturer04/21/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/21/2020
Type of Device Usage N
Patient Sequence Number1
Patient Age365 DA
-
-