• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNIVERS REVERS GLENO 36 +4 LAT; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERS REVERS GLENO 36 +4 LAT; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number UNIVERS REVERS GLENO 36 +4 LAT
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 03/20/2020
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event Description
It was reported that on (b)(6) 2019 a revers shoulder arthroplasty was performed.Everything went well and it was finished successfully.The surgeon was satisfied after the case.This was also showed on the x-rays.Approximately 6 months post-op the patient reported subsequently failing while putting on trousers.He felt an issue.This was when the glenoid side gave way.It was an elective revers and not a fracture revers.X-rays were done on (b)(6) 2020.A revision surgery is planned to be on (b)(6) 2020.Update 02-apr-2020: further information was received: the revision surgery went ahead on (b)(6) 2020 afternoon.It was revised with the new modular glenoid system, successfully using an iliac crest graft.Everything went very well.The failed devices were retrieved, however both peripheral screws were broken and the tips remain in the patient.Baseplate, central screw & remaining portion of peripheral screws x 2 will be returned assembled.Also from the humeral side the cup and liner, which needed to be removed for access, will be returned.There was no issue with the humeral side, however they will be returned regardless.
 
Manufacturer Narrative
Complaint not confirmed, no abnormality was observed on the device that may have contributed to the event.Additional information provided has been added to section b.
 
Event Description
***follow-up update on 29-apr-2020 : further information was provided that the the patient was elderly and the bone quality was not in top condition but also not too bad.The damage to the implants occurred (as the patient claims) when putting on trousers, this necessitated the revision, they did not break on removal, they were already broken from whatever trauma they suffered.It was further reported that it is very unlikely that just putting on trousers caused both screws to break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS REVERS GLENO 36 +4 LAT
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key9985419
MDR Text Key188864459
Report Number1220246-2020-01802
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888867061408
UDI-Public00888867061408
Combination Product (y/n)N
PMA/PMN Number
K142863
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model NumberUNIVERS REVERS GLENO 36 +4 LAT
Device Catalogue NumberAR-9504S-04
Device Lot Number18.01753
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/02/2020
Initial Date FDA Received04/22/2020
Supplement Dates Manufacturer Received04/02/2020
Supplement Dates FDA Received05/07/2020
Patient Sequence Number1
Patient Outcome(s) Other;
-
-