This report has been identified as b.Braun (b)(4) internal report # (b)(4).We only received one picture of a used perifix periduralkath.20g soft tip connected with a perifix cath.Con.2.1 out of a perifix-soft-tip 701 set 18g filter lor.Although the batch number was reported as unknown, the user facility did provide the following batch numbers that are in use at the facility: 19n02a8701 / 19m12a8701.The following investigations were conducted: visual inspection: the received picture was taken to a visual inspection for damages according to the test method.Definition of the method: damages are visible changes in the original shape or surface of products and packaging caused by mechanical forces and/or manufacturing faults.Nominal: no damage is allowed that endangers the patient, impedes the use of the part as intended (e.G.The impairment of the function of a drop sensor), endangers the assembly or function of the component or impairs the appearance of the component.Actual: the catheter is cut off, the cut off part of the catheter is not visible at the customer picture.Summary and assessment: we exclude a manufacturing fault since the catheters were taken to a 100% examination and therefore no mechanical damages or manufacturing faults are allowed.Because of this we assume of a problem during the application process.Based on the conducted investigations the checked sample is within the specification.Therefore the complaint is considered as not confirmed.The complaint is only taken to knowledge and filed for statistical purposes.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|