Model Number VTICM5_12.6
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Patient Involvement (2645)
Event Date 03/09/2020
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.No similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon was loading a 12.6mm vticm5_12.6 implantable collamer lens, -6.0/+1.5/006 (sphere/cylinder/axis), and found the lens damaged.There was no patient contact.The surgeon implanted a longer lens and the problem was resolved.
 
Manufacturer Narrative
Lens was returned in a micro-centrifuge vial, dry with clear surgical residue on the lens.Visual inspection found the lens haptic torn and residue on the lens.Claim#: (b)(4).
 
Manufacturer Narrative
Corrected data: b5.It was reported cause of the event was user error.Should have been included.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia, ca
MDR Report Key9988428
MDR Text Key191999600
Report Number2023826-2020-00952
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup,Followup
Report Date 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberVTICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2020
Initial Date Manufacturer Received 04/03/2020
Initial Date FDA Received04/22/2020
Supplement Dates Manufacturer Received07/21/2020
09/24/2020
Supplement Dates FDA Received07/27/2020
09/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARTRIDGE MODEL SFC-45 - LOT # UNK.; FOAM TIP PLUNGER MODEL FTP - LOT # UNK.; INJECTOR MODEL MSI-PF - LOT # UNK.
Patient Age22 YR
-
-