BLUE BELT TECHNOLOGIES ASSY, CPU, COMPUTER CART, NAVIO; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR220001
Device Problem Power Problem (3010)
Patient Problem Injury (2348)
Event Date 08/12/2016
Event Type  malfunction  
Event Description
It was reported that during a procedure, when advancing from the hip center screen, the screen showed some pixilation and checkered color patterns.After restarting the machine, the problem persisted.All monitor connections were double checked, but the behavior continued and the screen locked up and wouldn't advance by touch (pedals were not tried).The system was restarted multiple times with no success.The procedure was converted to manual.The investigation found the problem was the cpu, it had a defective/damaged graphics card.
 
Manufacturer Narrative
The device was used in treatment and it was reported that the monitor was displaying a pixelated view.Dhr review found that no conditions which could contribute to the reported event were identified.This information is reasonably suggesting that the device met the specifications at the date when it was released to the distribution.A complaint history found similar reports, this issue will continue to be monitored.The surgical system user's manual released at the time of the complaint provides instructions for troubleshooting computer issues.This is an identified failure mode within the risk assessment.We could confirm there was a relationship established between the reported event and the device.The device was returned for investigation.The returned monitor and cpu were tested and the complaint was confirmed.When the monitor with a known good cpu was tested, no problem was found.The cpu was tested with another known good monitor and the problem persisted.Swapped the graphics card on the cpu to rule out a graphics card issue and tested with the same "known good" monitor.The error did not present itself when attempted with a "known good" graphics card.The root cause of the error is most likely a defective/damaged graphics card.The root cause was supplier / raw material fault.A corrective action was opened for this issue.Per complaint details, the device malfunctioned during use and the navio was abandoned for manual instrumentation.Based on the information provided the impact beyond the reported modified procedure/conversion to manual could not be determined.No further medical assessment is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSY, CPU, COMPUTER CART, NAVIO
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
MDR Report Key9989223
MDR Text Key188648681
Report Number3010266064-2020-00277
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR220001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2016
Initial Date Manufacturer Received 08/12/2016
Initial Date FDA Received04/23/2020
Supplement Dates Manufacturer Received08/04/2020
Supplement Dates FDA Received08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age83 YR
-
-