Model Number 3100A
Device Problem Failure to Recalibrate (1517)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/10/2019
Event Type  malfunction  
Event Description
Oscillator switched out due to mean airway pressure (map) fluctuating from higher to lower than desired map.Multiple failed attempts to recalibrate the oscillator that the patient was originally on, so oscillator was switched out to a new one.The oscillator was pulled from the patient bedside and replaced with another vent.The oscillator was tagged and biomed was called to check it out.Biomed reviewed the oscillator and determined there was an issue with the piston within the unit.The parts have been ordered and will be replaced once in.The oscillator will be fixed and ensured it's operational before placed back in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100A
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC.
26125 north riverwoods blvd
mettawa IL 60045
MDR Report Key9990307
MDR Text Key188659108
Report Number9990307
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3100A
Device Catalogue Number768901
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/17/2020
Event Location Hospital
Date Report to Manufacturer04/23/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/23/2020
Type of Device Usage N
Patient Sequence Number1
Patient Age365 DA
-
-