• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P RSP 22.5IN-LB TORQUE LIMITING DRIVER; DRIVER, PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P RSP 22.5IN-LB TORQUE LIMITING DRIVER; DRIVER, PROSTHESIS Back to Search Results
Catalog Number 804-06-009
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Instrument failure - 30 minute delay in surgery.
 
Additional Manufacturer Narrative
See h4 the agent reported "this was a primary surgery.2 different torque screw drivers broke the set screw in the glenoid.We ended up wasting.5 implants because of this.Female 59 yrs." this event occurred during surgery, near the patient.No response was received by the surgeon.The surgery was completed as intended, with a thirty-minute delay.The instrument was inspected prior to use and was deemed acceptable for use based on its appearance.The agent was present during surgery and was able to source a suitable replacement device.Rma examination: the instruments were returned to djo and after further examination, the tips of the torque driver are stripped.None of the implants came back with the instruments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP 22.5IN-LB TORQUE LIMITING DRIVER
Common Device Name
DRIVER, PROSTHESIS
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key17981431
Report Number1644408-2023-01530
Device Sequence Number11767387
Product Code HWR
UDI-Device Identifier00190446187899
UDI-Public00190446187899
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/26/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number804-06-009
Device Lot Number234456L11 & 300110L20
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2023
Supplement Date Received by Manufacturer11/27/2023
Initial Report FDA Received Date10/20/2023
Supplement Report FDA Received Date12/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-03-126 LOT: 834C2162.; 508-00-001 LOT: 850C1396C.; 508-32-101 LOT: 862C4188.; 508-32-204 LOT: 769P2841.; 508-32-204 LOT: 769P2858.
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
Patient SexFemale
-
-