• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH 4.5MM PERIPHERAL SCREW - 40MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH 4.5MM PERIPHERAL SCREW - 40MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number 5572-4540
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was reported that a patient underwent had their shoulder stem removed 6 weeks post opp due to metallosis.
 
Additional Manufacturer Narrative
The reported event could not be confirmed.The device inspection revealed the following: the lot # of the returned 4.5mm peripheral screw - 40mm could not be identified.The screw is relatively new, no major wear mark was visible.It was not possible to determine if metallic particles were created.It was not possible to confirm the reported metallosis, as not enough information was provided in this case.Furthermore, as indicated by medical experts, "metallosis is about the type-iv immune response to the accumulation of metal particles within tissues.There are cases of early arthroplasty failure due to metallosis, yet these patients usually do well initially and start to get pain and joint stiffness after months, leading to revision within 2 years.".However, the reported metallosis was observed after 6 weeks of implantation, a post-surgical lapse of time much shorter than what is usually observed for such response.This gives further indication of the lack of evidence pointing towards metallosis and the impossibility to confirm the reported event.Based on the investigation, no problem could be detected on the screw and the link with a possible metallosis could not be established.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that a patient underwent had their shoulder stem removed 6 weeks post opp due to metallosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM PERIPHERAL SCREW - 40MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15883197
Report Number0008031020-2022-00658
Device Sequence Number12074750
Product Code KWS
UDI-Device Identifier07613327098778
UDI-Public07613327098778
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number5572-4540
Device Catalogue Number5572-4540
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/23/2022
Initial Report FDA Received Date11/29/2022
Supplement Report FDA Received Date01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-