• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS, INC. COMBO INSERTER/IMPACTOR. S

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS, INC. COMBO INSERTER/IMPACTOR. S Back to Search Results
Model Number GLN-0926-030-01A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.Case-2022-00131670-1.
 
Event or Problem Description
It was reported that, during surgery, a fin-lock glenoid, s, machined broke during the inserting of the implant.Surgery was finished without any delay with the same device.No harm to the patient or further complications reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
B5 (describe event or problem).
 
Event or Problem Description
It was reported that, during surgery, a combo inserter/impactor.S broke during the inserting of the implant.Surgery was finished without any delay with the same device.No harm to the patient or further complications reported.
 
Additional Manufacturer Narrative
D4: catalog number and unique identifier (udi) #.
 
Additional Manufacturer Narrative
H3, h6: the product has not been received at the aus site for evaluation and photographs were not provided, so the reported event could not be confirmed.The following investigative actions were performed.Complaint history review: the complaint history review identified similar reported events for this device and for the same lot.No adverse trend was identified.Future complaints for this failure mode will continue to be monitored and investigated as required.Risk management review (device): identified no issues which could have caused or contributed to the reported event.The failure mode was previously identified by the risk management file and the anticipated risk level is acceptable.Dhr/batch record review: identified no issues or manufacturing abnormalities which could have caused or contributed to the reported event.This review determined that the device met manufacturing specifications upon release for distribution.Capa/nc/pra/hhe/field action review: identified no previous events or issues which could have caused or contributed to the reported event.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Device labeling/ifu review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met labeling requirements upon release for distribution.Product prints/specifications/procedure review: identified no issues which could have caused or contributed to the reported event.This review determined that the device met product specifications upon release for distribution.A definitive root cause could not be determined.The complaint alleges no injury to the patient nor a delay during surgery.A determination of whether the device might have contributed to the reported event could not be made as a definitive root cause for the failure could not be determined.The fin-lock combo inserter/impactor is a reusable instrument expected to be used across multiple surgeries.During normal use, the inserter grips the fin-lock glenoid implant and secures it to the glenoid impactor handle.It places the glenoid implant in the correct location and then protects the implant while the handle is being impacted with a mallet to seat the implant.Therefore, the inserter may fail if excessive force is applied during impaction or may weaken over time due to normal wear.Additionally, due to the design of the instrument, the inserter may be broken if mishandled, such as by resting a heavy object on top of the device while the bottom is unsupported.Based on this investigation, the need for corrective action is not indicated as no non-conformances or manufacturing deficiencies were identified and the risk level is acceptable.If additional information is later received, the complaint may be reopened.No further investigation is warranted for this complaint, though future complaints will be monitored and investigated as required.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.The visual inspection revealed that a piece of the reusable instrument was fractured off.Additional signs of wear were identified along the edges of the devices, specifically indents and scratches.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMBO INSERTER/IMPACTOR. S
Common Device Name
IMPACTOR
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16015258
Report Number3002788818-2022-00188
Device Sequence Number12355227
Product Code HWA
UDI-Device Identifier00885556860731
UDI-Public885556860731
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K152047
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 02/14/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGLN-0926-030-01A
Device Catalogue NumberIMP-0926-030-01A
Device Lot Number208065-02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/23/2022
Supplement Date Received by Manufacturer01/13/2023
01/20/2023
01/20/2023
02/12/2024
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date01/17/2023
01/23/2023
01/24/2023
02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-