SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 52MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 74122552
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, after undergoing left hip bhr surgery on (b)(6) 2009 in which an unknown bhr cup and modular femoral head were implanted, the patient experienced unspecified complications.This adverse event was treated by performing a revision surgery on (b)(6) 2022.The patient subsequently presented additional unspecified adverse events that were treated by performing posterior capsule mesh reconstruction on (b)(6) 2022 (documented under case-2023-00142670).
 
Additional Manufacturer Narrative
H6: medical device problem code section h3, h6:it was reported that a left hip revision surgery was performed due to unspecified metal-related complications.As of today, the implanted devices, all of which were used in treatment are not accessible for testing.As no device part and batch numbers were provided for investigation for any of the 3 devices, a complaint history review, manufacturing record review, or definite escalation actions review could not be performed.If more information is received, this investigation will be reopened.The review of the product's current ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed for the reported metal-related complications.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions for all hemi heads and sleeves and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.As no device part and batch numbers were provided for the cup, a escalation actions review could not be performed.The available medical documents were reviewed.Based on the reported symptoms it cannot be concluded that the reported event/clinical reaction (unspecified metal-related complications) were associated with a mal-performance of the implant.The patient impact cannot be determined at this time.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Additional Manufacturer Narrative
B4 - describe event or problem - it was reported that, after undergoing left hip bhr-tha surgery on (b)(6) 2009, the patient experienced unspecified metal-related complications.This adverse event was treated by performing a revision surgery on (b)(6) 2022.Additionally, on (b)(6) 2022, the patient underwent a posterior capsule mesh reconstruction to repair extensive damage to the hip.D1 - brand name ¿ unknown birmingham hip modular head (bhmh) hemi.D2a product code ¿ kwl.D2b common device name - prosthesis, hip, hemi-, femoral, metal.
 
Event or Problem Description
It was reported that, after undergoing left hip bhr-tha surgery on (b)(6) 2009, the patient experienced unspecified metal-related complications.This adverse event was treated by performing a revision surgery on (b)(6) 2022.Additionally, on (b)(6) 2022, the patient underwent a posterior capsule mesh reconstruction to repair extensive damage to the hip.
 
Event or Problem Description
It was reported that, after undergoing left hip bhr-tha surgery on (b)(6) 2009, the patient experienced unspecified metal-related complications.This adverse event was treated by performing a revision surgery on (b)(6) 2022.
 
Additional Manufacturer Narrative
H10: additional information was added in b5, d1, d4, d10, and h4.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, after undergoing left hip bhr-tha surgery, the patient experienced unspecified metal-related complications.As of today, the devices, which were used in treatment, have not been returned for evaluation.A review of the historical complaints data for the hemi head, modular sleeve and bhr cup was performed, using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe of 12 months prior to the date in which the incident was reported.Similar complaints have been identified and this will continue to be monitored through routine monthly trending, it should be noted that the sleeve and head are no longer sold.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.A review of historic quality escalations was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible; no further escalation actions required.The available medical documents were reviewed.The reported increasing metal levels and pain, pseudotumor, and corrosion material and debris around the taper and in the taper sleeve may be consistent with the reported adverse local reaction.However, the clinical root cause of the reported events/clinical reactions cannot be confirmed, and it cannot be concluded that the reported events/clinical reactions were associated with a mal performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Based on the information provided we cannot further investigate the complaint our investigation remains inconclusive and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.If the products or additional information become available in the future, this case will be reopened.Based on this investigation the need for corrective and preventative action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 52MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16342786
Report Number3005975929-2023-00006
Device Sequence Number12463437
Product Code KWL
UDI-Device Identifier00885556071694
UDI-Public885556071694
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/20/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2013
Device Catalogue Number74122552
Device Lot Number08LW20009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2023
Supplement Date Received by Manufacturer01/24/2023
03/22/2023
05/01/2023
07/20/2023
Initial Report FDA Received Date02/09/2023
Supplement Report FDA Received Date02/20/2023
03/24/2023
05/25/2023
07/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71356006 / LOT# 09LM04941; 74122158 / LOT# 08GW17648; 74222300 / LOT# 09HW24565
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-