• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BIA300 IMPLANT 4MM W ABUTMENT 9MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BIA300 IMPLANT 4MM W ABUTMENT 9MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 92346
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Unspecified Infection (1930); Inadequate Osseointegration (2646)
Type of Reportable Event Serious Injury
Event or Problem Description
Per the clinic, the patient experienced infection at the implant site and a loss of osseointegration resulting in fixture loss.The patient was treated with oral antibiotics (specific date and duration not reported).Additional information has been requested but has not been made available as of the date of this report.
 
Additional Manufacturer Narrative
This report is submitted on november 20, 2023.
 
Additional Manufacturer Narrative
Per the clinic, the patient underwent skin revision surgery to excise the excess skin (specific date not reported).The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.Implant loss with baha implants can occur at any time following implantation.Known reasons for implant loss include lack of adequate bone quality/quantity, trauma, infection, generalised diseases and surgical complication.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on december 26, 2023.
 
Additional Manufacturer Narrative
Per the clinic, the patient underwent silver nitrate cauterization in order to remove excess skin (specific date not reported).This report is submitted on march 19, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA300 IMPLANT 4MM W ABUTMENT 9MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
nabila zahari
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18167902
Report Number6000034-2023-03933
Device Sequence Number12953525
Product Code MAH
UDI-Device Identifier09321502019767
UDI-Public(01)09321502019767(10)1563387(17)270718
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K100360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/21/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number92346
Device Catalogue Number92346
Device Lot NumberCOH1563387
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer11/29/2023
02/21/2024
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date12/26/2023
03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;