• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE SHOULDER; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ10X108MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE SHOULDER; ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ10X108MM Back to Search Results
Model Number 508-01-432
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/07/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the stem being loose, making the shoulder unstable.The stem was removed.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to loosening.The previous surgery and the revision detailed in this investigation occurred over 1 year and 9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Additional Manufacturer Narrative
D10 - adding concomitant part 402283.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE SHOULDER
Common Device Name
ENCORE REVERSE SHOULDER HUMERAL STEM, STD, SZ10X108MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key7281175
Report Number1644408-2018-00141
Device Sequence Number1354961
Product Code PHX
UDI-Device Identifier00888912024723
UDI-Public(01)00888912024723
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K141990
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/27/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/18/2022
Device Model Number508-01-432
Device Catalogue Number530-10-108
Device Lot Number409T1044
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/07/2018
Supplement Date Received by Manufacturer02/23/2018
02/23/2018
Initial Report FDA Received Date02/19/2018
Supplement Report FDA Received Date02/27/2018
05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
402283 COBALT G-HV BONE CEMENT 40G; 508-01-432, LOT 922F1905; 530-10-108, LOT 409T1044
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-