|
DEPUY SYNTHES PRODUCTS LLC 1.8MM FLUTED MATCHSTICK, 15CM; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Model Number |
MA15-8N-18 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during cleaning it was observed that the unknown burr device was stuck in the attachment device and broke off inside the attachment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Concomitant medical devices and therapy dates: attachment device, (b)(6) 2020.Brand name was unknown.Udi: gtin was unknown.Catalog number, serial number/lot number are unknown.The manufacturing location was unknown.510(k) number was unknown.Device manufacture date was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.Quality engineering evaluated the cutter device, and the reported condition that the device was broken off and stuck inside the attachment device was confirmed.The device was taken apart and it was noted that some corrosion, and medical debris was in the locking mechanism assembly and was preventing the lock tabs from releasing the cutters which caused the failure.The piece of the cutter was examined using 25x magnification and rechecked on an optical comparator.A full assessment of the device could not be performed due to the condition of the cutter was received.The piece of the cutter was broken off below the laser marking and identification of the cutter and the lot number was not able to be identified and the rest of the cutter was not sent in.The cause for cutter getting stuck was likely due to debris and residue preventing the lock tabs from moving into place.The cause of the medical debris and detergent residue buildup was due to improper cleaning (user error).The assignable root cause of this condition was determined to be traced to maintenance, which is improper maintenance.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device has been returned and is currently pending evaluation.Once the evaluation of the device has been completed, a supplemental medwatch report will be sent accordingly.G4: please note that the brand name, common device name, device product code, unique identifier (udi) and 510k number have been received and have been entered into this supplemental medwatch report.
|
| |
|
Search Alerts/Recalls
|
|
|