• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 115-222
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Failure of Implant (1924); Seroma (2069)
Date of Event 11/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported right side "rupture." the device has been explanted and replaced.
 
Event or Problem Description
Patient reported diagnostic findings for right side as "fluid collection", "internal and external capsular rupture", and "a malignant neoplasm of unknown breast".
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b.5, b.6, h.6.The event of seroma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed a broken sharp assessed as unidentified (tear) opening (shell thickness was within specification).Seroma-late: unable to confirm as it is a medical event not related to the device.Additional observations: crease flat observed in the device.Wear abrasion observed in the device.No further actions are required as the device was implanted.
 
Event or Problem Description
Patient reported right side "rupture." patient reported diagnostic findings for right side as "fluid collection", "internal and external capsular rupture", and "a malignant neoplasm of unknown breast".The device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Clarification to b.5 and h.6: it had not been confirmed that surgery occurred as scheduled on 18nov2022 in previous medwatch submission.
 
Event or Problem Description
Patient reported right side "rupture." the device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15917683
Report Number9617229-2022-22056
Device Sequence Number14574561
Product Code FTR
UDI-Device Identifier10888628002340
UDI-Public10888628002340
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 03/09/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/07/2022
Device Model Number115-222
Device Catalogue Number115-222
Device Lot Number3016565
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer12/09/2022
12/29/2022
02/16/2023
Initial Report FDA Received Date12/05/2022
Supplement Report FDA Received Date12/29/2022
01/12/2023
03/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-