• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 40MM M SPEC+15.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 40MM M SPEC+15.5; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-09-000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Pocket Erosion (2013); Loss of Range of Motion (2032); Discomfort (2330); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Date of Event 10/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
(b)(6) claim letter record received.Claim letter alleges pain, discomfort, difficulty walking, metallosis, permanent tissue and muscle loss, reduced ability to work, loss of a normal life, and elevated metal ions.However, lab results shows below 7 ppb.Mri suggest adverse local tissue reaction.The patient is seeking compensation for all the injuries experienced.Doi: (b)(6) 2011; dor: (b)(6) 2021 ; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Through follow up it is now understood there is no allegation associated against a product malfunction.Review of the x-rays evidence found nothing indicative of a device nonconformance.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
 
Event or Problem Description
On (b)(6) 2020, progress notes reported that the patient experienced left knee pain and hip pain.About 2 weeks ago patient felt popping in left hip area and constant pain in external rotation.On (b)(6) 2021 patient had a broken left hip prosthesis.On (b)(6) 2021, patient undergone left hip revision.No operative notes provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 clinical symptoms code: unspecified tissue injury (e2015) used to capture soft tissue injury and bone injury.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
After review of the medical records the patient was revised to address failed tha secondary to adverse tissue reaction associated with a mom prosthesis, pseudotumor, elevated metal ions, metal wear resulting to hip pain and discomfort.X-rays reported significant erosion of the bone along the medial calcar.Showed some bone loss on medial calcar.Operative note reported large brownish gray fluid.Tissues were collected, dark brown consistent with an adverse tissue reaction.There was some blackening of the trunnion but no corrosion.Hip was reduced with a fair amount of effort.Before there was signs of subluxation of the head.Lab result reported below 7 ppb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 40MM M SPEC+15.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds IN LS11 8DT
UK   LS11 8DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13838605
Report Number1818910-2022-05035
Device Sequence Number14688154
Product Code LPH
UDI-Device Identifier10603295033042
UDI-Public10603295033042
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/18/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2012
Device Model Number1365-09-000
Device Catalogue Number136509000
Device Lot Number2482951
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer03/22/2022
01/12/2023
03/16/2023
04/07/2023
Initial Report FDA Received Date03/21/2022
Supplement Report FDA Received Date03/23/2022
01/19/2023
03/17/2023
04/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/25/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINN CAN BONE SCREW 6.5MMX25MM; PINN SECTOR W/GRIPTION 56MM; PINNACLE MTL INS NEUT40IDX56OD; TRI-LOCK BPS SZ 5 HI OFFSET; UNK HIP ACETABULAR CUP PINNACLE; UNK HIP ACETABULAR LINER METAL PINNACLE; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight164 KG
Patient RaceWhite
-
-