• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 UKNOWN NASOGASTRIC TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 UKNOWN NASOGASTRIC TUBE Back to Search Results
Model Number EN0066000
Medical Device Problem Codes Component Missing (2306); Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the customer were experienced difficulties in removing suction tubing from the nasogastric tube once it has been connected.Customer also stated that they had bile and secretions came out of the antireflux valve end of the tube which was getting on patients and bedding.Customer asked is there some sort of positioning that they need to do to help with this problem.Per additional information received via email on 07apr2022, customer stated that the enfit tubes did not have the prevent filter attached.Stated that it was the little disc shaped integral plug component at the end of the pigtail.Inquirer responded that the plug was just for when they were transporting the patient and not sure why the enfit version did not include it.Inquirer assumed that there was an engineering reason but could not said for sure, it should not have any impact on the issue.
 
Event or Problem Description
It was reported that the customer were experienced difficulties in removing suction tubing from the nasogastric tube once it has been connected.Customer also stated that they had bile and secretions came out of the antireflux valve end of the tube which was getting on patients and bedding.Customer asked is there some sort of positioning that they need to do to help with this problem.Per additional information received via email on (b)(6)2022, customer stated that the enfit tubes did not have the prevent filter attached.Stated that it was the little disc shaped integral plug component at the end of the pigtail.Inquirer responded that the plug was just for when they were transporting the patient and not sure why the enfit version did not include it.Inquirer assumed that there was an engineering reason but could not said for sure, it should not have any impact on the issue.
 
Additional Manufacturer Narrative
The reported event is inconclusive.It is unknown whether the device had met relevant specifications.It is unknown whether the product had caused the reported failure.Visual evaluation of the attached photo sample noted 1 opened (without original packaging) bard suction connector connected to suction tubing.Visual inspection noted no obvious visible defects.The reported event cannot be confirmed without the physical sample received.Further investigation cannot be conducted.A potential root cause could be incorrect fixture.The dhr review could not be performed without a lot number.The product catalog number for this device is unknown.Therefore, bd is unable to determine the associated labeling to review.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was evaluated.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the customer were experienced difficulties in removing suction tubing from the nasogastric tube once it has been connected.Customer also stated that they had bile and secretions came out of the antireflux valve end of the tube which was getting on patients and bedding.Customer asked is there some sort of positioning that they need to do to help with this problem.Per additional information received via email on 07apr2022, customer stated that the enfit tubes did not have the prevent filter attached.Stated that it was the little disc shaped integral plug component at the end of the pigtail.Inquirer responded that the plug was just for when they were transporting the patient and not sure why the enfit version did not include it.Inquirer assumed that there was an engineering reason but could not said for sure, it should not have any impact on the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UKNOWN NASOGASTRIC TUBE
Common Device Name
NASOGASTRIC TUBE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14282030
Report Number1018233-2022-03262
Device Sequence Number14837970
Product Code BSS
UDI-Device Identifier00801741128707
UDI-Public(01)00801741128707
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/19/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEN0066000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer09/23/2022
10/20/2022
Initial Report FDA Received Date05/04/2022
Supplement Report FDA Received Date10/05/2022
10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-