DEXCOM, INC. DEXCOM G6 PROFESSIONAL SYSTEM; CONTINUOUS GLUCOSE MONITOR Back to Search Results
Catalog Number STP-PR-001
Medical Device Problem Code Premature End-of-Life Indicator (1480)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that early sensor expiration occurred.The sensor was inserted into the abdomen no product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
Cmpl-(b)(4).
 
Event or Problem Description
It was reported that early sensor expiration occurred.The sensor was inserted into the abdomen.Data was evaluated and the allegation was confirmed.The probable cause was determined to be high counts aberration via data.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 PROFESSIONAL SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key15494507
Report Number3004753838-2022-178936
Device Sequence Number14994134
Product Code QII
UDI-Device Identifier00386270001092
UDI-Public00386270001092
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K191833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberSTP-PR-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date09/27/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-