MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BI300 IMPLANT 3MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BI300 IMPLANT 3MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 92128
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Skin Infection (4544)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on july 11, 2023.
 
Event or Problem Description
Per the clinic, the patient experienced infection at the implant site.Subsequently, the device was explanted (date not reported).There are no plans to reimplant the patient with a new device, as of the date of this report.
 
Additional Manufacturer Narrative
Per the clinic, the patient was placed under general anesthesia on (b)(6) 2023.The patient was also treated with oral antibiotics (specific date and duration not reported).This report is submitted on august 11, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI300 IMPLANT 3MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
thevamalar jayamagentharan
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key17296454
Report Number6000034-2023-02143
Device Sequence Number15049845
Product Code LXB
UDI-Device Identifier09321502019538
UDI-Public(01)09321502019538(10)COH1311015(17)240522
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K100360
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/22/2024
Device Model Number92128
Device Catalogue Number92128
Device Lot NumberCOH1311015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/15/2023
Supplement Date Received by Manufacturer08/07/2023
Initial Report FDA Received Date07/11/2023
Supplement Report FDA Received Date08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-