• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Cyst(s) (1800); Failure of Implant (1924)
Date of Event 01/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial knee arthroplasty on an unknown date.Subsequently, the patient was revised due to a loose tibia with a cyst under the medial side.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: development of rounded lucency along the medial and central portion of the tibial component most consistent with osteolysis.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING LEFT SIZE F
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18490889
Report Number0001822565-2024-00142
Device Sequence Number15218062
Product Code MBH
UDI-Device Identifier00889024510579
UDI-Public(01)00889024510579(17)320315(10)65314693
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K173417
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42530007501
Device Lot Number65314693
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2023
Supplement Date Received by Manufacturer01/15/2024
05/23/2024
Initial Report FDA Received Date01/11/2024
Supplement Report FDA Received Date01/16/2024
05/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight102 KG
-
-