• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS; SEE H10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS; SEE H10 Back to Search Results
Model Number SPECIFIC DEVICE NOT REPORTED
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377); Metal Related Pathology (4530); Insufficient Information (4580)
Date of Event 04/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
As reported via legal documentation the patient had a left hip replacement on (b)(6) 2015.Approximately 8 years and 2 months after the initial procedure the patient had a left hip replacement on (b)(6) 2023.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Event or Problem Description
Revision operative report on (b)(6) 2023-pre-operative long diagnosis: failed left total hip replacement secondary to poly wear.Evidence of osteolysis and poly wear involving mainly the acetabulum.The poly was noted to be wearing excessively.There was noted ostial lysis behind the screw holes as well as osteolysis over the superior rim of the acetabular component.Femoral head removed; there was slight trunnion metallosis.Bone scan negative for loosening.Revision left tha of acetabular component only.The patient tolerated the procedure well.No other information is provided.
 
Additional Manufacturer Narrative
Additional information: h6 health effect - clinical code, medical device problem code.D2a.Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented d10.Concomitants: 3601984 120-65-20 - bone screw 6.5mm dia x 20mm long.2816164 160-31-16 - pf spline plasma w/ha ext offset sz 16.2419988 170-40-00 - biolox delta femoral 40mm od, +0m.3807630 186-01-58 - integrip cc, cluster 58mm.These devices are used for treatment not diagnosis.Pending investigation.No other information is available.
 
Additional Manufacturer Narrative
The most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this cannot be not confirmed with the information provided; devices were not returned.There is no other information available.These devices are used for treatment not diagnosis.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6: clinical codes, component code, type of investigation, investigation findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER LIPPED 40MM ID, GROUP 3 CUPS
Common Device Name
SEE H10
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
michael crader
3523771140
MDR Report Key17490561
Report Number1038671-2023-01915
Device Sequence Number15409377
Product Code LZO
UDI-Device Identifier10885862207043
UDI-Public10885862207043
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/28/2026
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2019
Device Model NumberSPECIFIC DEVICE NOT REPORTED
Device Catalogue Number132-40-53
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer09/06/2023
02/21/2024
04/14/2026
Initial Report FDA Received Date08/09/2023
Supplement Report FDA Received Date09/27/2023
03/18/2024
04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/19/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1732-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-