• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE CD; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE CD; INSTRUMENT, KNEE Back to Search Results
Catalog Number 42517000303
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an initial total knee procedure, the tibial articular surface provisional instrument was chipped and unable to be inserted during the trialing process.There was no patient impact and no adverse events have been reported as a result of the malfunction.It was reported that no further information is available.
 
Additional Manufacturer Narrative
(b)(4).D10: medical product: mc tibial articular surface provisional left size 4-5 cd top: catalog#42517100310, lot#ni.Multiple mdr reports have been filed for this report; please see associated report: 0001822565-2023-03419.The product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.H3 other text : see h10 narrative.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: suggested component code: mechanical (g04) - provisional bottom.The reported event was confirmed via provided photographs.Visual examination of the pictures identified the unit as chipped.A review of the device history records could not be performed as lot number was not provided.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d1; d2; d4; d9; d10; g3; g4; h2; h3; h4; h6; h10.D10: medical product: mc tibial articular surface provisional left size 4-5 cd top: catalog#42517100310, lot#65120752 the product has been received by zimmer biomet and prompted a review of the previously reported investigation results.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, d9, g3, g6, h2, h6, h11.The complaint sample was returned, and visual examination of the product identified that it exhibited signs of repeated use and was worn, rather than fractured as originally reported.The root cause remains unchanged at this time.As the device was received worn, and no event of articular surface provisional wear has resulted in serious injury, zimmer biomet does not consider this event to be reportable.Additionally, upon review of our reporting decision on fracture of articular surface provisionals, a review of all complaints for this product indicates that there has never been an event that has resulted in a death or serious injury.Therefore, zimmer biomet will not be reporting this failure mode going forward.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE CD
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18256014
Report Number0001822565-2023-03418
Device Sequence Number15715751
Product Code MBH
UDI-Device Identifier00889024238459
UDI-Public(01)00889024238459(10)65158058
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/29/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42517000303
Device Lot Number65158058
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer12/29/2023
02/20/2024
07/02/2024
Initial Report FDA Received Date12/04/2023
Supplement Report FDA Received Date01/02/2024
03/14/2024
07/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.; SEE H10 NARRATIVE.
Patient SexPrefer Not To Disclose
-
-