MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIOINDUCTIVE IMPLANT W/ARTH DEL LRG; MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIOINDUCTIVE IMPLANT W/ARTH DEL LRG; MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON Back to Search Results
Model Number 4566
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Pain (1994)
Date of Event 09/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that, approx.6 months post-operation, the patient returned to the clinic reporting pain and inflammation.The patient said a mattress hot his shoulder.The mris showed inflammation and the presence of rice bodies around the repair site.A recent mri looked good and normal, rice bodies and inflammation were gone.Patient had an infection and recovered following antibiotics.No follow-up surgery/washout was scheduled.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).Corrected information in d1, d2, d4 (catalog & udi), and g4 (510k).
 
Additional Manufacturer Narrative
H10: h2: additional information: h6: health effect - impact code.H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined.Due to the limited clinical information provided.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that no relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Based on the limited information provided a thorough medical assessment could not be performed; therefore, we are unable to conclude specific factors known to contribute to the alleged report.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOINDUCTIVE IMPLANT W/ARTH DEL LRG
Common Device Name
MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11823502
Report Number3003604053-2021-00188
Device Sequence Number16415501
Product Code OWY
UDI-Device Identifier00885556735664
UDI-Public885556735664
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K140300
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/25/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/22/2023
Device Model Number4566
Device Catalogue Number4566
Device Lot NumberA7954
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer11/04/2022
02/23/2024
Initial Report FDA Received Date05/13/2021
Supplement Report FDA Received Date11/04/2022
02/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-