• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV MEDICAL TECHNOLOGIES CHAMELEON; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV MEDICAL TECHNOLOGIES CHAMELEON; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number CH08-40-75US
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Open or Remove Packaging Material (2922); Device Fell (4014)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, there was no damage noted to the balloon, but when the balloon was prepared, and the protective sheath was pulled off of the catheter prior to the procedure or prior to using on the patient, the balloon detached from the catheter shaft; the product was no longer used.The physician used another product (entire unit) from the same manufacturer to complete the procedure.There was no damage to the device's box and packaging.The product/kit was still sealed upon receipt.There was no excessive force used to pull/take out the product.There was nothing unusual observed on the device prior to use, and flushing was not done.There were no other products being utilized with the device, and there were no other defects/damages found on the product.There was no blood loss, and blood transfusion was not required.There was no intervention/treatment required as a result of the event.There was no patient injury.
 
Event or Problem Description
According to the reporter, there was no damage noted to the balloon, but when the balloon was prepared and the protective sheath was pulled off of the catheter prior to the procedure or prior to using on the patient, the balloon detached from the catheter shaft.It was mentioned that the detachment was spontaneous, and no excessive force was used to remove the protective sheath, which caused the balloon to detach.The product was no longer used.The physician used another product (entire unit) from the same manufacturer to complete the procedure.There was no damage to the device's box and packaging.The product/kit was still sealed upon receipt.There was no excessive force used to pull/take out the product.There was nothing unusual observed on the device prior to use, and flushing was not done.There were no other products being utilized with the device, and there were no other defects/damages found on the product.There was no blood loss, and blood transfusion was not required.There was no intervention/treatment required as a result of the event.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: b5, d8, d9, g3, h3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the balloon was detached from the catheter shaft and was not returned.Microscopic inspection of the distal end of the catheter shaft noted that the shaft tubing was torn directly after the infusion plate.It was reported that there was detachment of the balloon.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHAMELEON
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
AV MEDICAL TECHNOLOGIES
21 habarzel street
tel-aviv 67770 16
Manufacturer (Section G)
AV MEDICAL TECHNOLOGIES
21 habarzel street
tel-aviv 67770 16
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key10731679
Report Number3011144059-2020-00007
Device Sequence Number16732108
Product Code KRA
UDI-Device Identifier07290016745047
UDI-Public07290016745047
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K170635
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/01/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberCH08-40-75US
Device Catalogue NumberCH08-40-75US
Device Lot Number21903676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2020
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer11/20/2020
07/02/2021
06/28/2022
06/28/2022
02/21/2023
Initial Report FDA Received Date10/24/2020
Supplement Report FDA Received Date12/14/2020
07/08/2021
06/28/2022
07/01/2022
03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-