• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. NORM-JECT-F LUER SOLO 1 ML (NJ-9166017-02); SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. NORM-JECT-F LUER SOLO 1 ML (NJ-9166017-02); SYRINGE, PISTON Back to Search Results
Model Number NJ-9166017-02
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
We are a 503b outsourcing facility.After repackaging of an avastin lot, we observed two blue particles within finished product syringes.A long investigation ensued and we were able to determine the lot of syringes have particles in it from the manufacturer.We had to reject the lot of avastin we made due to this.We placed the lot of syringes on qa hold and ordered a new lot.The new lot also contains particles.The particles are only visible under magnification; therefore, we inspected sealed syringes under 15x magnification before making another lot (which was also sent for subvisible particle testing), but feel as through these syringes are entirely unsafe to continue to be used for any type of injection.We have ceased further production using these syringes, which accounts for over 60% of our business.We reached out to the manufacturer (air-tite / b.Braun) and they have started a complaint/investigation as well (reference #(b)(4)).We have inspected unused and sealed syringes from three (3) separate lots, and all have been found to have varying levels of particles.Item description norm-ject-f luer solo 1 ml (nj-9166017-02) lot # dom, exp qty inspected, qty with fod 21b11c8, 2021-02-11, 2026-02-01, 222 14 20m23c8; 2020-11-23, 2025-11-01, 200 3 21d26c8, 2021-04-26, 2026-04-01 100, 8.Inspection at 15x magnification recorded in section, "tell us what happened and how it happened." air-tite and b.Braun; device only problem, no patient impact at this point.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NORM-JECT-F LUER SOLO 1 ML (NJ-9166017-02)
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
MDR Report Key12407214
Report NumberMW5103652
Device Sequence Number17040407
Product Code FMF
Initial Reporter StateID
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Voluntary
Initial Reporter Occupation Unknown
Type of Report Initial
Report Date (Section B) 08/30/2021
3 Devices were Involved in the Event: 1   2   3  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberNJ-9166017-02
Device Catalogue NumberNJ-9166017-02
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age1 DA
-
-