• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. STAY SAFE MTS SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. STAY SAFE MTS SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 050-50758
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the multi stay safe connector during dwell 2 of 6 of their pd treatment.The cause of the leak is unknown.The patient was assisted in changing the flow alert option to no.Upon follow up, it was confirmed the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the multi stay safe connector during dwell 2 of 6 of their pd treatment.The cause of the leak is unknown.The patient was assisted in changing the flow alert option to no.Upon follow up, it was confirmed the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the multi stay safe connector during dwell 2 of 6 of their pd treatment.The cause of the leak is unknown.The patient was assisted in changing the flow alert option to no.Upon follow up, it was confirmed the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the multi stay safe connector during dwell 2 of 6 of their pd treatment.The cause of the leak is unknown.The patient was assisted in changing the flow alert option to no.Upon follow up, it was confirmed the patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy with no further issues.The cycler set used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Correction provided in d1, d2, d4, g4, and h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAY SAFE MTS SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key18894382
Report Number0008030665-2024-00264
Device Sequence Number17113740
Product Code KDI
UDI-Device Identifier00840861100699
UDI-Public00840861100699
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K173651
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/08/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2025
Device Catalogue Number050-50758
Device Lot Number22NR08006
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2024
Supplement Date Received by Manufacturer03/22/2024
03/28/2024
04/03/2024
Initial Report FDA Received Date03/13/2024
Supplement Report FDA Received Date03/27/2024
04/01/2024
04/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; DELFLEX PD FLUID.; DELFLEX PD FLUID.; DELFLEX PD FLUID.; LIBERTY CYCLER SET.; LIBERTY CYCLER SET.; LIBERTY CYCLER SET.; LIBERTY CYCLER SET.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Patient SexMale
-
-