Model Number 1060657
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Event or Problem Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
The manufacturer previously reported hospitalization - initial or prolonged in b.2 section, which was marked incorrectly as this is product problem only as mentioned in b.1 section in previous report.A.1 section was inorrect patient intial name was reported in the previous report, which correctly updated in this follow up report as unknown.G.2 section was missing the previous report, in this correction report it is correctly marked as consumer.E.1 section was missing in the previous report, which is correctly updated in this follow up report.D.9 and h.3 section incorrectly reported as no in the previous report, which is correctly reported as yes in this follow up report.H.10 comment was missing in previous report, the correct comment is now updated in follow up report as: this mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.Type of investigation in h.6 also missed to mention 10(testing of actual/suspected device), which correctly updated now.D.1 and d.2 sections were wrongly reported in the previous report, which are correctly updated in this follow up report.Pma/510(k) number in g.4 wrongly updated in previous report, which is correctly updated as k102465 in this follow up report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AVAPS
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13594748
Report Number2518422-2022-06000
Device Sequence Number17440088
Product Code MNS
UDI-Device Identifier00606959043985
UDI-Public00606959043985
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1060657
Device Catalogue Number1060657
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 12/04/2014
Supplement Date Received by Manufacturer12/04/2014
Initial Report FDA Received Date02/24/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-